Are you sure you want to leave this community? Leaving the community will revoke any permissions you have been granted in this community.
Etienne Laliberté, Raymond Soffer 2018. Measuring spectral reflectance and transmittance (350-2500 nm) of small and/or narrow leaves using the Spectra Vista Corporation (SVC) DC-R/T Integrating Sphere. protocols.io https://dx.doi.org/10.17504/protocols.io.qp7dvrnCopy Citation Copied
URL: https://dx.doi.org/DOI:10.17504/protocols.io.qp7dvrn
Authors: Etienne Laliberté, Raymond Soffer
Group: Canadian Airborne Biodiversity Observatory
Summary: Here we describe the standardised protocol used by the Canadian Airborne Biodiversity Observatory (CABO) to measure leaf spectral reflectance and transmittance, using an integrating sphere fitted to a portable full-range field spectroradiometer, for the special case where an individual leaf is too small and/or too narrow to entirely cover the reflectance or transmission port of the integrating sphere. Briefly, three arrays of mature, healthy and sunlit leaves from a canopy plant are arranged on a custom sample mount, and are then used for measurements of adaxial reflectance and transmittance. Reflectance measurements are referenced to a calibrated Spectralon® disk and corrected for stray light. Our leaf spectroscopy protocol is an adaptation of that of Noda et al. (2013) to the SVC DC-R/T sphere, and also builds from protocols by the Carnegie Airborne Observatory and the SVC integrating sphere user manual.
Affiliations: Université de Montréal, National Research Council
Version: 2
Publication Date: 2018
Expand AllComing soon.
Source: