Ariel Rabines, Rob Lampe, Andrew Allen 2020. Sterivex DNA extraction. protocols.io https://dx.doi.org/10.17504/protocols.io.bc2hiyb6Copy Citation Copied
URL: https://dx.doi.org/DOI:10.17504/protocols.io.bc2hiyb6
Authors: Ariel Rabines, Rob Lampe, Andrew Allen
Group: A.E. Allen Lab
Summary: A mostly automated protocol for extraction of genomic DNA from seawater filtered onto a Sterivex filter (Cat. No. SVGP0150). Reagents come from the Macherey-Nagel NucleoMag Plant Kit for DNA purfication (Cat. No. 744400). Automated liquid handling is performed on an eppendorf EpMotion 5075t with multi-channel pipettes. This protocol assumes that the Sterivex is sealed with tube sealant on the male end and a male-luer lock plug on the female end. Using this protocol, we routinely extract ~80 samples at a time.The protocol has been modified to increase to an 800 µL starting volume. This normally results in less Lysis Buffer MC1 and Binding Buffer MC2 than is needed. Extra binding buffer should be purchased separately. The kit manual can be found here: https://www.mn-net.com/media/pdf/09/46/a0/Instruction-NucleoMag-Plant.pdfAdditional notes from the manufacturer on using the kit with the epMotion can be found here: https://www.mn-net.com/media/pdf/95/2c/ef/AN-NucleoMag-Plant-epMotion5075.pdf . Note that we have modifications to their protocol.The epMotion program is attached here.
Affiliations: Scripps Institution of Oceanography, Scripps Institution of Oceanography, Scripps Institution of Oceanography
Version: 2
Publication Date: 2020